Input Voltage |
Voltage | 115 / 230VAC ± 15%, User selection |
Frequency | 50 / 60 Hz ± 5% |
Fuse | 2A / 250V Slow-Blow fuse |
Line condition |
Power Switch | Reverse polarity switch setting select ON / OFF |
Neutral Switch | Neutral switch on/off selection for single fault condition |
Ground Switch | Ground switch on/off selection for class I single fault condition |
Probe Setting |
Ground to Line (G – L) | V | V |
Surface to Line (PH – L) | V | - |
Surface to Surface (PH – PL) | V | - |
Ground to Neutral (G – N) | - | V |
Auto Function (G - L & G - N) | - | V |
SETTING |
Leakage HI / Lo Trip | Range | 0uA, 1uA - 10000uA ( 0 = OFF ) | 0uA, 1uA - 10000uA ( 0 = OFF ) |
Resolution | 1uA | 1uA |
LeakageHI / Lo Trip (Peak) | Range | 0uA, 1uA - 15000uA ( 0 = OFF) | - |
Resolution | 1uA |
Leakage Current RMS Display |
Range | MD Major Resistance is 0.5KΩ | MD Major Resistance is 1KΩ | MD Major Resistance is 1.5KΩ |
Range 1 | 0.0uA - 260.0uA | 0.0uA - 130.0uA | 0.0uA - 85.0uA |
Range 2 | 240.0uA - 999.9uA, 1000uA - 1050uA | 120.0uA - 525.0uA | 80.0uA - 350.0uA |
Range 3 | 800.0uA - 999.9uA, 1000 - 4200uA | 400.0uA - 999.9uA, 1000uA - 2100uA | 266.6uA - 999.9uA, 1000uA - 1400uA |
Range 4 | 3600uA - 10000uA | 1800uA - 10000uA | 1200uA - 10000uA |
Leakage Current RMS Display |
Range | Frequency | Basic Accuracy |
Range 1 - 4 (for ESC-112) | DC, 15Hz - 10KHZ | ± (2% of reading + 5 counts) |
10K-1MHZ | ± (5% of reading) |
Range 1 - 4 (for ESC-125) | DC, 15Hz - 200KHZ | ± (2% of reading + 5 counts) |
200K - 1MHZ | ± (5% of reading) |
Leakage Current Peak Display ( for ESC-112 ) |
Range | MD Major Resistance is 0.5KΩ | MD Major Resistance is 1KΩ | MD Major Resistance is 1.5KΩ |
Range 1 | 0uA - 260uA | 0uA - 130uA | 0uA - 85uA |
Range 2 | 240uA - 1050uA | 120uA - 525uA | 80uA - 350uA |
Range 3 | 800uA - 4200uA | 400uA - 2100uA | 266uA - 1400uA |
Range 4 | 3600uA - 15000uA | 1800uA - 15000uA | 1200uA - 15000uA |
Leakage Current Peak Display ( for ESC-112 ) |
Range | Frequency | Basic Accuracy |
Range 1 - 4 | DC, 15Hz - 1MHZ | ± (10% of reading + 2uA) |
MD Circuit module |
Measuring Device (MD) | A, B, C, D, E, F, G | A, E, F, G |
MD A. UL 544 Non Patient, UL484 |
MD B. UL 544 Patient Care |
MD C. IEC60601-1, UL2601-1, EN 60601-1 |
MD D. UL1563 |
MD E. UL60950, IEC60950, IEC61010 - 1, IEC60335-1, IEC60990 |
MD F. External MD |
MD G. Basic measuring element 1KΩ of Frequency check |
MD Components | Resistance accuracy = 1%; Capacitance accuracy = 5% |
MD Voltage Limit | Maximum 32Vdc |
Internal Leakage | 1. Internal Leakage current = 65uA, 2. 277V applied to Ph max leakage current = 70uA |
Leakage Current Offset | Range : 0 - 999.9uA ; Resolution : 0.1uA |
DUT POWER |
AC Voltage Display | Range | 30.0- 300.0V |
Resolution | 0.1V |
Accuracy | ± (1% of reading + 2 counts) , 30.0 - 300.0VAC |
AC Current | 30A max continuous |
Over current protection | 32A, Response time < 600ms / 68Apeak, Response time < 600ms |
Delay Time | Range | 0, 1.0 - 999.9s (0 = continuous) |
Resolution | 0.1s |
GENERAL |
PLC Remote Control | Input : Test, Reset, and Recall Memory 1 through 3, Output : Pass, Fail, Processing for ESC-112; Pass, Fail, Processing, Start-Out, Reset-Out for ESC-125 |
Memory | 99 memories, 8step / memory |
Display | 20 x 2 OLED |
Alarm Volume Setting | Range : 0 - 9; 0 = OFF, 1 is softest volume, 9 is loudest volume |
Key Lock | To prevent unauthorized alteration of the test Parameters |
Calibration | Software and adjustments are made through front panel |
Interface | Optional USB & RS232, GPIB |
Environment | 0 - 40ºC, 20 - 80%RH |
Dimension (W xH x D), mm | 430 x 89 x 400 |
Net Weight | 11.5Kg |
*product specifications are subject to change without notice. |
Input Voltage |
Voltage | 115 / 230VAC ± 15%, User selection |
Frequency | 50 / 60 Hz ± 5% |
Fuse | 2A / 250V Slow-Blow fuse |
Line condition |
Power Switch | Reverse polarity switch setting select ON / OFF |
Neutral Switch | Neutral switch on/off selection for single fault condition |
Ground Switch | Ground switch on/off selection for class I single fault condition |
Probe Setting |
Ground to Line (G – L) | V | V |
Surface to Line (PH – L) | V | - |
Surface to Surface (PH – PL) | V | - |
Ground to Neutral (G – N) | - | V |
Auto Function (G - L & G - N) | - | V |
SETTING |
Leakage HI / Lo Trip | Range | 0uA, 1uA - 10000uA ( 0 = OFF ) | 0uA, 1uA - 10000uA ( 0 = OFF ) |
Resolution | 1uA | 1uA |
LeakageHI / Lo Trip (Peak) | Range | 0uA, 1uA - 15000uA ( 0 = OFF) | - |
Resolution | 1uA |
Leakage Current RMS Display |
Range | MD Major Resistance is 0.5KΩ | MD Major Resistance is 1KΩ | MD Major Resistance is 1.5KΩ |
Range 1 | 0.0uA - 260.0uA | 0.0uA - 130.0uA | 0.0uA - 85.0uA |
Range 2 | 240.0uA - 999.9uA, 1000uA - 1050uA | 120.0uA - 525.0uA | 80.0uA - 350.0uA |
Range 3 | 800.0uA - 999.9uA, 1000 - 4200uA | 400.0uA - 999.9uA, 1000uA - 2100uA | 266.6uA - 999.9uA, 1000uA - 1400uA |
Range 4 | 3600uA - 10000uA | 1800uA - 10000uA | 1200uA - 10000uA |
Leakage Current RMS Display |
Range | Frequency | Basic Accuracy |
Range 1 - 4 (for ESC-112) | DC, 15Hz - 10KHZ | ± (2% of reading + 5 counts) |
10K-1MHZ | ± (5% of reading) |
Range 1 - 4 (for ESC-125) | DC, 15Hz - 200KHZ | ± (2% of reading + 5 counts) |
200K - 1MHZ | ± (5% of reading) |
Leakage Current Peak Display ( for ESC-112 ) |
Range | MD Major Resistance is 0.5KΩ | MD Major Resistance is 1KΩ | MD Major Resistance is 1.5KΩ |
Range 1 | 0uA - 260uA | 0uA - 130uA | 0uA - 85uA |
Range 2 | 240uA - 1050uA | 120uA - 525uA | 80uA - 350uA |
Range 3 | 800uA - 4200uA | 400uA - 2100uA | 266uA - 1400uA |
Range 4 | 3600uA - 15000uA | 1800uA - 15000uA | 1200uA - 15000uA |
Leakage Current Peak Display ( for ESC-112 ) |
Range | Frequency | Basic Accuracy |
Range 1 - 4 | DC, 15Hz - 1MHZ | ± (10% of reading + 2uA) |
MD Circuit module |
Measuring Device (MD) | A, B, C, D, E, F, G | A, E, F, G |
MD A. UL 544 Non Patient, UL484 |
MD B. UL 544 Patient Care |
MD C. IEC60601-1, UL2601-1, EN 60601-1 |
MD D. UL1563 |
MD E. UL60950, IEC60950, IEC61010 - 1, IEC60335-1, IEC60990 |
MD F. External MD |
MD G. Basic measuring element 1KΩ of Frequency check |
MD Components | Resistance accuracy = 1%; Capacitance accuracy = 5% |
MD Voltage Limit | Maximum 32Vdc |
Internal Leakage | 1. Internal Leakage current = 65uA, 2. 277V applied to Ph max leakage current = 70uA |
Leakage Current Offset | Range : 0 - 999.9uA ; Resolution : 0.1uA |
DUT POWER |
AC Voltage Display | Range | 30.0- 300.0V |
Resolution | 0.1V |
Accuracy | ± (1% of reading + 2 counts) , 30.0 - 300.0VAC |
AC Current | 30A max continuous |
Over current protection | 32A, Response time < 600ms / 68Apeak, Response time < 600ms |
Delay Time | Range | 0, 1.0 - 999.9s (0 = continuous) |
Resolution | 0.1s |
GENERAL |
PLC Remote Control | Input : Test, Reset, and Recall Memory 1 through 3, Output : Pass, Fail, Processing for ESC-112; Pass, Fail, Processing, Start-Out, Reset-Out for ESC-125 |
Memory | 99 memories, 8step / memory |
Display | 20 x 2 OLED |
Alarm Volume Setting | Range : 0 - 9; 0 = OFF, 1 is softest volume, 9 is loudest volume |
Key Lock | To prevent unauthorized alteration of the test Parameters |
Calibration | Software and adjustments are made through front panel |
Interface | Optional USB & RS232, GPIB |
Environment | 0 - 40ºC, 20 - 80%RH |
Dimension (W xH x D), mm | 430 x 89 x 400 |
Net Weight | 11.5Kg |
*product specifications are subject to change without notice. |