泰克電流探頭CT6

具體成交價以合同協議為準

聯系我們時請說明是化工儀器網上看到的信息,謝謝!


 

Features & Benefits

  • High Bandwidth
  • Ultra-low Inductance
  • Very Small Form Factor
  • Characterize Current Waveforms up to <200 ps Rise Times
  • Very Low Loading of Circuit Under Test
  • Fits Into Dense, Closely-spaced Circuit Designs

Applications

  • Data Storage Read Channel Design
  • Silicon Characterization
  • High-frequency Analog Design
  • ESD Testing
  • Signal Injection
  • Differential Current Measurements
  • Single-shot Low Rep-rate Pulse Measurements
  • Propagation Delay Measurement

泰克電流探頭CT6Current Probe

The CT6 is the newest addition to the Tektronix portfolio of high-frequency current probes. It is designed to meet the needs of high-speed circuit design and test applications which require ultra-high bandwidth, low inductance, and extremely small form factor. The CT6 provides up to 2 GHz bandwidth when used with high-bandwidth oscilloscopes such as the Tektronix TDS694C, TDS794D, and TDS7000 Series oscilloscopes or with other compatible 50 Ω input measuring instruments. Low inductance (<3 nH) assures that the loading effect of the CT6 on the circuit-under-test will be negligible, which is especially important for today's low-amplitude, high-speed circuit designs such as disk drive read/write preamplifiers. The probe is a closed-circuit design which will accept uninsulated wire sized up to 20 gauge. This product is exempt from CE mark by virtue of its 30 V voltage limit.

CT1/CT2 Current Probes

The CT1 and CT2 Current Probes are designed for permanent or semi-permanent in-circuit installation. Each probe consists of a current transformer and an interconnecting cable. The current transformers have a small hole through which a current carrying conductor is passed during circuit assembly.

The P6041 Probe Cable provides the connection between the CT1 and CT2 Current Transformers and a BNC oscilloscope input. A 50 Ω termination is required to terminate the cable when connected to a high-impedance (1 MΩ) oscilloscope input. One probe cable can be used to monitor several current transformers that have been wired into a circuit.


泰克CT1電流探頭/CT2. Current Probes with P6041 BNC Probe Cable.

MINIATURE CONSTRUCTION

泰克CT1電流探頭 and CT2 detachable cable design enables one or more probes to be located on circuit boards or in other limited space areas.

The CT6 offers the smallest form factor available, for measurement on ever-shrinking circuit boards and components. It is designed for temporary installation and does not incorporate removable cables, as the CT1 and CT2 do.

EXTENDIBLE PROBE LENGTH

Specified rise time and bandwidth are obtained when using the probe cables provided: The P6041 cable used with 泰克CT1電流探頭and CT2 is 42 inches nominal. If additional length is required, the cables can be extended by using high-quality 50 Ω cable and suitable interface connectors. (Also see Special Probe Cables, Optional Accessories.) Long cables may degrade high-frequency response.

HIGH SENSITIVITY

The CT1 and泰克電流探頭CT6 provide an output of 5 mV for each milliamp of input current when terminated in 50 Ω. The CT2 provides 1 mV per milliamp when terminated in 50 Ω.

TYPICAL SYSTEMS

The CT1, CT2, and high-frequency current transformers are dynamic (i.e., non-DC) current measuring devices. They are typically used in conjunction with compatible high-bandwidth oscilloscopes and other instruments to observe and/or record high-frequency current waveforms. The CT1, CT2, and CT6 normally operate directly into 50 Ω scopes and other measuring device inputs.

The CT1 or CT2 can be used with 1 MΩ input systems; use the P6041 probe cable and terminate the output with a 50 Ω feed-through termination (see Optional Accessories).

In all cases, the CT1, CT2, and CT6 must work into 50 Ωs to obtain specified performance and sensitivity.

Typical Measurement Applications

DIFFERENTIAL CURRENT MEASUREMENTS

Most true-differential voltage amplifiers have a maximum bandwidth of about 100 MHz. The CT1 or CT6 can make differential current measurements to 1 GHz and 2 GHz, respectively, by passing two wires carrying opposing currents through the same core. The displayed result is the difference current. The CT2 can perform the same function to 200 MHz.

In all cases, Derating with Frequency and Amp-second Product (Current-time Product) guidelines should not be exceeded. (See Characteristics.)

SINGLE-SHOT AND LOW REP-RATE PULSE MEASUREMENTS

These common measurements are easy to make with the CT1, CT2, or CT6 provided that your signal fits within the Max Pulse Current and Amp-second Product (Current-time Product) guidelines for the specific current probe characteristics.

For example, the CT2 is rated at 36 A peak, with an Amp-second Product of 50 × 10-6 seconds (50 Amp-microseconds), therefore the CT2 can safely handle a 36 A peak pulse with a maximum width of 1.39 microseconds or lower amplitude pulses for longer pulse widths. The CT1, CT2, and CT6 all have low-frequency roll-off characteristics. Low-frequency "droop" will exhibit itself when the pulse width approaches the L/R time constant of the specific transformer.

PROPAGATION DELAY MEASUREMENTS

Two CT1 or CT2 Current Transformers with matching probe cables can be used to measure propagation delay (transit time) between the input and output currents of high-frequency devices. The probe outputs are connected to the inputs of dual-channel real-time or sampling scopes.

Verification of any Probe/Cable/Scope System mismatch can be obtained by passing the same signal current through both probes and observing total system delay difference, if any.

 

CT6 Typical Frequency Response.

CT1, CT2, and CT6 Characteristics

Characteristic

CT1

CT2

CT6

Bandwidth (typical)

25 kHz to 1 GHz

1.2 kHz to 200 MHz

250 kHz to 2 GHz

Rise Time

350 ps

500 ps

200 ps

Sensitivity

(into 50 Ω)

5 mV/mA

1 mV/mA

5 mV/mA

Accuracy

±3%

±3%

±3%

Magnetizing Inductance

6 μH

7 μH

1 μH

Leakage Inductance

2.4 nH

1 nH

1.5 nH

Insertion Impedance:

at 10 MHz

<1 Ω

0.1 Ω

1.1 Ω

at 100 MHz

2 Ω

0.5 Ω

1.3 Ω

at 1 GHz

 

 

11.9 Ω

Max. Bare Wire Size

#14 wire

#16 wire

#20 wire

1.78 mm (0.070 in.)

1.32 mm (0.052 in.)

0.8 mm (0.032 in.)

Max. Bare Wire Voltage:

RMS

175 VRMS CAT I

175 VRMS CAT I

30 VRMSCAT I

Peak

1000 V*1

1000 V*1

30 V

DC

175 V

175 V

30 V

Max. Peak Pulse Current

12 A

36 A

6 A

Max. Continuous Current (RMS)

450 mA

2.5 A

120 mA

Amp-second Product

1 × 10-6 A*Sec

50 × 10-6 A*Sec

0.25 × 10-6 A*Sec

L/R Time Constant (droop)

>6.35 μs

>160 μs

0.4 μs

Propagation Delay

5.4 ns

6.1 ns

5.2 ns

Safety

UL3111-2-032, CSA1010.2.032, EN61010-2-032, IEC61010-2-032

UL3111-2-032, CSA1010.2.032, EN61010-2-032, IEC61010-2-032

NA

*1 (<3.25% Duty Factor).

 

 


 

 

 

Ordering Information

CT1

High-frequency Current Probe.

Includes: Manual and P6041 Interconnect Cable.

CT2

High-frequency Current Probe.

Includes: Manual and P6041 Interconnect Cable.

CT6

High-frequency Current Probe.

 

SERVICE

Opt. C3 - Calibration Service 3 Years

Opt. C5 - Calibration Service 5 Years

Opt. D1 - Calibration Data Report (CT6 only)

Opt. D3 - Calibration Data Report 3 Years (with Opt. C3)

Opt. D5 - Calibration Data Report 5 Years (with Opt. C5)

Opt. R3 - Repair Service 3 Years

Opt. R5 - Repair Service 5 Years

Opt. SILV400 - Standard Warranty Extended to 5 Years (CT6 Only).

 

CT1 OR CT2 RECOMMENDED ACCESSORIES

Feed-through 50 Ω Termination - Order 011-0049-02.

 

CT6 STANDARD ACCESSORIES

SMA to BNC Adapter - Order 015-0572-xx.

User Manual - Order 071-0453-xx.

Color Cable Marker Bands -

Probe Holder - Order 015-0682-xx.

Certificate of Traceable Calibration -

Warranty - One Year.



化工儀器網

采購商登錄
記住賬號    找回密碼
沒有賬號?免費注冊

提示

×

*您想獲取產品的資料:

以上可多選,勾選其他,可自行輸入要求

個人信息:

溫馨提示

該企業已關閉在線交流功能

主站蜘蛛池模板: 久久狠狠色情网| 欧美激情欧美激情| 国产欧美一区二区三区免费| 又硬又粗进去好爽A片免费视频| 啊灬啊灬快灬高潮了视频| 酒色婷婷| 亚洲一区精品在线| 麻豆文化传媒网站地址| 国产日韩欧美二区| 欧美日韩激情一区二区三区| 亚洲天堂av手机天堂网| 金妍儿不雅视频| 久久久这里只有精品免费| 神马手机观看限制电影| 黑料网-独家爆料破解版| 观看欧美日韩一级片| 激情五月开心婷婷深爱| 精品久久亚洲一区| 国产无遮挡又黄又爽免费网站| 国产午夜精品福利久久| 亚洲VA欧美VA天堂V国产综合| 16影院理论片在线观看| 国产瑜伽白皙一区二区| 久草热8精品视频在线观看| 日韩熟妇中文字幕| 一级女人毛片人一女人| 亚洲永久无码7777| 久久久久久久久麻豆| 久久久久久久久免费视频| 国产69精品久久久| 男女爽爽午夜18污污影院| 亚洲精品久久| 学生媚薬痉挛中文字幕| 日韩中文字幕在线视频观看| 717韩国三级理伦电影| 亚洲中文字幕永久在线全国| 欧美韩国日本一区| 黄色免费久久久久| 高清毛片AAAAAAAAA片| 国产乱理伦片在线观看网站| 亚洲最大成人网站|